Mechanical and geometric properties of thermoformed and 3D printed clear dental aligners

      Highlights

      • Successful modeling and 3D printing of clear dental aligners.
      • Signify the effects of postcuring 3D printed clear dental aligners.
      • Higher compression strength of 3D printed than thermoformed clear dental aligners.
      • Greater geometric accuracy of 3D printed than thermoformed clear dental aligners.
      • Lesser processing costs of 3D printed than thermoformed clear dental aligners.

      Introduction

      The aim of this research was to compare compressive mechanical properties and geometric inaccuracies between conventionally manufactured thermoformed Duran clear dental aligners and 3D printed Dental Long Term (LT) resin-based clear aligners using 3D modeling and printing techniques.

      Methods

      Impressions of the patient's dentition were scanned and using 3D modeling software, dental models were designed and 3D printed. These printed models then underwent vacuum thermoforming to thermoform a clear Duran thermoplastic sheet of 0.75-mm thickness into clear dental aligners of the same thickness of 0.75 mm. For the same dental model, aligners were also designed and 3D printed to 0.75-mm thickness creating biocompatible clear dental aligners using Dental LT resin utilizing a Formlabs 3D printing machine for direct usage by the patients. Five observers calculated teeth height for both types of aligners for evaluation of geometric deviations. Both types of aligners were subjected to compression loading of 1000 N to evaluate their load vs displacement behavior.

      Results

      3D printed cured clear dental aligners were found to be geometrically more accurate with an average relative difference in tooth height of 2.55% in comparison with thermoformed aligners (4.41%). Low standard deviations (0.03-0.09 mm) were observed for tooth height measurements taken by all the observers for both types of aligners. 3D printed aligners could resist a maximum load of nearly 662 N for a low displacement of 2.93 mm; whereas, thermoformed aligners could resist a load on only 105 N for 2.93-mm displacement. Thermoformed aligners deformed plastically and irreversibly for large displacements; whereas, 3D printed aligners deformed elastically with reversibility for lower displacements.

      Conclusions

      3D printed and suitably cured Dental LT resin-based clear dental aligners are suggested to be more suitable for patient use as they are geometrically more accurate; this presents an opportunity to make processing time savings while ensuring an aligner is mechanically stronger and elastic in comparison with the conventionally produced thermoplastic-based thermoformed clear dental aligners.
      To read this article in full you will need to make a payment

      References

        • Birkeland K.
        • Katle A.
        • Løvgreen S.
        • Bøe O.E.
        • Wisth P.J.
        Factors influencing the decision about orthodontic treatment. A longitudinal study among 11- and 15-year-olds and their parents.
        J Orofac Orthop. 1999; 60: 292-307
        • Mueller B.
        Additive manufacturing technologies—rapid prototyping to direct digital manufacturing.
        Assem Autom. 2012; 32
        • Levy G.N.
        • Schindel R.
        • Kruth J.P.
        Rapid manufacturing and rapid tooling With layer manufacturing (LM) technologies, state of the art and future perspectives.
        CIRP Ann. 2003; 52: 589-609
        • Yan X.
        • Gu P.
        A review of rapid prototyping technologies and systems.
        Comput Des. 1996; 28: 307-318
        • Hazeveld A.
        • Huddleston Slater J.J.R.
        • Ren Y.
        Accuracy and reproducibility of dental replica models reconstructed by different rapid prototyping techniques.
        Am J Orthod Dentofacial Orthop. 2014; 145: 108-115
        • El-Katatny I.
        • Masood S.H.
        • Morsi Y.S.
        Error analysis of FDM fabricated medical replicas.
        Rapid Prototyp J. 2010; 16: 36-43
        • Lee K.Y.
        • Cho J.W.
        • Chang N.Y.
        • Chae J.M.
        • Kang K.H.
        • Kim S.C.
        • et al.
        Accuracy of three-dimensional printing for manufacturing replica teeth.
        Korean J Orthod. 2015; 45: 217-225
        • Bajaj D.
        • Madhav I.
        • Juneja M.
        • Tuli R.
        • Jindal P.
        Methodology for stress measurement by transparent dental aligners using strain gauge.
        World J Dent. 2018; 9: 13-18
        • Malik O.
        • McMullin A.
        • Waring D.
        Invisible orthodontics part 1.
        Invisalign. 2013; (203-4, 207-10, 213-215)
        • Barone S.
        • Paoli A.
        • Neri P.
        • Razionale A.V.
        • Giannese M.
        Mechanical and geometrical properties assessment of thermoplastic materials for biomedical application.
        in: Eynard B. Nigrelli V. Oliveri S.M. Peris-Fajarnes G. Rizzuti S. Advances on Mechanics, Design Engineering and Manufacturing : Proceedings of the International Joint Conference on Mechanics, Design Engineering & Advanced Manufacturing (JCM 2016), September 14-16, 2016, Catania, Italy Cham. Springer International Publishing, Berlin2017: 437-446
        • Ahn H.W.
        • Kim K.A.
        • Kim S.H.
        A new type of clear orthodontic retainer incorporating multi-layer hybrid materials.
        Korean J Orthod. 2015; 45: 268-272
        • Johal A.
        • Sharma N.R.
        • McLaughlin K.
        • Zou L.F.
        The reliability of thermoform retainers: A laboratory-based comparative study.
        Eur J Orthod. 2015; 37: 503-507
        • Nasef A.A.
        • El-Beialy A.R.
        • Mostafa Y.A.
        Virtual techniques for designing and fabricating a retainer.
        Am J Orthod Dentofacial Orthop. 2014; 146: 394-398
        • Lombardo L.
        • Martines E.
        • Mazzanti V.
        • Arreghini A.
        • Mollica F.
        • Siciliani G.
        Stress relaxation properties of four orthodontic aligner materials: A 24-hour in vitro study.
        Angle Orthod. 2017; 87: 11-18
        • Kohda N.
        • Iijima M.
        • Muguruma T.
        • Brantley W.A.
        • Ahluwalia K.S.
        • Mizoguchi I.
        Effects of mechanical properties of thermoplastic materials on the initial force of thermoplastic appliances.
        Angle Orthod. 2013; 83: 476-483
        • Juneja M.
        • Thakur N.
        • Kumar D.
        • Gupta A.
        • Bajwa B.
        • Jindal P.
        Accuracy in dental surgical guide fabrication using different 3-D printing techniques.
        Addit Manuf. 2018; 22: 243-255
        • Ryu J.H.
        • Kwon J.S.
        • Jiang H.B.
        • Cha J.Y.
        • Kim K.M.
        Effects of thermoforming on the physical and mechanical properties of thermoplastic materials for transparent orthodontic aligners.
        Korean J Orthod. 2018; 48: 316-325
        • Zguris Z.
        How mechanical properties of stereolithography 3D prints are affected by UV curing.
        Formlabs White Paper. 2016;
      1. Formlabs. EU declaration of conformity; 2017. Available at: https://formlabs.com/media/upload/Declaration_of_Conformity_Dental_LT_Clear_DoC-FLOC-2017-01-UK.pdf. Accessed February 18, 2019.

      2. Formlabs. Dental LT clear: Class IIa long term biocompatible resinfor form 2; 2017. Available at: https://formlabs.com/media/upload/DentalLTClear-DataSheet-EN.pdf. Accessed February 18, 2019.

      3. Formlabs. Printing splints with dental LT clear resin; 2017. Available at: https://support.formlabs.com/s/article/Printing-Splints-with-Dental-LT-Clear-Resin?language5en_US. Accessed February 18, 2019.

      4. Formlabs. Form wash and form cure tech specs; 2017. Available at: https://formlabs.com/tools/wash-cure/tech-specs. Accessed February 18, 2019.

      5. Formlabs. Form cure time and temperature settings; 2017. Available at: https://support.formlabs.com/s/article/Form-Cure-Timeand-Temperature-Settings?language5en_US. Accessed February 18, 2019.

        • Kitching I.
        Direct manufacture of orthodontic aligner appliance.
        2013
        • Hague R.
        • Mansour S.
        • Saleh N.
        • Harris R.
        Materials analysis of stereolithography resins for use in Rapid Manufacturing.
        J Mater Sci. 2004; 39: 2457-2464
        • Loza-Herrero M.A.
        • Rueggeberg F.A.
        • Caughman W.F.
        • Schuster G.S.
        • Lefebvre C.A.
        • Gardner F.M.
        Effect of heating delay on conversion and strength of a post-cured resin composite.
        J Dent Res. 1998; 77: 426-431
        • Kim K.
        • Choy K.
        • Park Y.C.
        • Han S.Y.
        • Jung H.
        • Choi Y.J.
        Prediction of mandibular movement and its center of rotation for nonsurgical correction of anterior open bite via maxillary molar intrusion.
        Angle Orthod. 2018; 88: 538-544