Force delivery properties of thermoplastic orthodontic materials

  • Jae-Sung Kwon
    Affiliations
    Graduate student, Department of Dental Biomaterial Science, College of Dentistry, Seoul National University, Seoul, Korea.
    Search for articles by this author
  • Yong-Keun Lee
    Correspondence
    Reprint requests to: Yong-Keun Lee, Department of Dental Biomaterial Science, College of Dentistry, Seoul National University, 28 Yeongeon-dong, Jongro-gu, Seoul, Korea
    Affiliations
    Professor, Department of Dental Biomaterial Science and Dental Research Institute, College of Dentistry, Seoul National University, Seoul, Korea.
    Search for articles by this author
  • Bum-Soon Lim
    Affiliations
    Professor, Department of Dental Biomaterial Science, College of Dentistry, Seoul National University, Seoul, Korea.
    Search for articles by this author
  • Yong-Kyu Lim
    Affiliations
    Associate professor, Department of Orthodontics, Graduate School of Clinical Dentistry, Korea University, Seoul, Korea.
    Search for articles by this author
      Introduction: Our objectives were to evaluate the force and energy (resilience) delivery properties of thermoplastic overlay orthodontic materials and to determine the changes in force delivery properties after thermocycling or repeated load cycling. Methods: Three types and 3 thicknesses of materials were investigated. Three-point bending-recovery tests were performed at baseline and after cycling procedures. For cycling, the specimens were thermocycled for 1000 cycles or repeatedly deflected by 1 mm for 100 cycles. Vickers hardness was measured to determine the changes after thermocycling or repeated load cycling. Results: The amount of deflection for optimal force delivery was 0.2 to 0.5 mm. Thin material exerted high energy in the deflection range of optimal force delivery. In the deflection ranges of optimal force delivery (0.2-0.5 mm), the force delivery properties after thermocycling were not different from those at the baseline (P >.01) but were different after repeated load cycling (P <.01). Thermocycling and repeated load cycling influenced Vickers hardness significantly. Conclusions: Thin material (0.508 mm) can deliver higher energy than thick materials (0.762 or 1.016 mm, P <.01) of the same brand. Therefore, thin material should be selected in the same brand of material. The effect of repeated deflection during service should be considered.
      To read this article in full you will need to make a payment

      References

        • Ponitz R.J.
        Invisible retainers.
        Am J Orthod. 1971; 59: 266-272
        • McNamara J.A.
        • Kramer K.L.
        • Juenker J.P.
        Invisible retainers.
        J Clin Orthod. 1985; 19: 570-578
        • Sheridan J.J.
        • Ledoux W.
        • McMinn R.
        Essix retainers: fabrication and supervision for permanent retention.
        J Clin Orthod. 1993; 27: 37-45
        • Sheridan J.J.
        • Armbruster J.
        Bleaching teeth during supervised retention.
        J Clin Orthod. 1999; 33: 339-344
        • Rinchuse D.J.
        • Rinchuse D.J.
        Active tooth movement with Essix-based appliances.
        J Clin Orthod. 1997; 31: 109-112
        • Kesling H.D.
        The philosophy of the tooth positioning appliance.
        Am J Orthod. 1945; 31: 297-304
        • Boyd R.L.
        • Miller R.J.
        • Vlaskalic V.
        The Invisalign system in adult orthodontics: mild crowding and space closure case.
        J Clin Orthod. 2000; 34: 203-212
        • Sheridan J.J.
        • Ledoux W.
        • McMinn R.
        Essix appliance: minor tooth movement with divots and windows.
        J Clin Orthod. 1994; 28: 659-663
        • Burstone C.J.
        The biophysics of bone remodeling during orthodontics—optimal force considerations.
        in: Norton L.A. Burstone C.J. The biology of tooth movement. CRC Press, Boca Raton, La1989: 321-334
        • Proffit W.R.
        Contemporary orthodontics.
        C. V. Mosby, St Louis2000
        • Sheridan J.J.
        • Hilliard K.
        • Armbruster P.
        Essix appliance technology: applications, fabrication and rationale.
        GAC International, Bohemia, NY2003
        • Waked E.J.
        • Lee T.K.
        • Caputo A.A.
        Effects of aging on the dimensional stability of custom-made mouthguards.
        Quintessence Int. 2002; 33: 700-705
        • McNamara J.A.
        • Brudon W.L.
        Orthodontic and orthopedic treatment in the mixed dentition.
        in: Needham Press, Ann Arbor, Mich1993: 347-353
        • Lindauer S.J.
        • Shoff R.C.
        Comparison of Essix and Hawley retainers.
        J Clin Orthod. 1998; 32: 95-97
        • Schuster S.
        • Eliades G.
        • Zinelis S.
        • Eliades T.
        • Bradley T.G.
        Structural conformation and leaching from in vitro aged and retrieved Invisalign appliances.
        Am J Orthod Dentofacial Orthop. 2004; 126: 725-728
        • Craig R.G.
        • Powers J.M.
        Restorative dental materials.
        in: 11th ed. Mosby, St Louis2002: 102
        • Chenin D.A.
        • Trosien A.H.
        • Fong P.F.
        • Miller R.A.
        • Lee R.S.
        Orthodontic treatment with a series of removable appliances.
        J Am Dent Assoc. 2003; 134: 1232-1239
        • Vlaskalic V.
        • Boyd R.
        Orthodontic treatment of a mildly crowded malocclusion using the Invisalign system.
        Aust Orthod J. 2001; 17: 41-46
        • Kayalioglu M.
        • Toroglu M.S.
        • Uzel I.
        Tooth-size ratio for patients requiring 4 first premolar extractions.
        Am J Orthod Dentofacial Orthop. 2005; 128: 78-86
        • Begg P.R.
        Differential force method of orthodontic treatment.
        Am J Orthod. 1956; 42: 481-510
        • Storey E.
        • Smith R.
        Force in orthodontics and its relation to tooth movement.
        Aust Dent J. 1952; 56: 11-18
        • Lee B.W.
        Relationship between tooth-movement rate and estimated pressure applied.
        J Dent Res. 1965; 44: 1053
        • Lee B.W.
        The force requirements for tooth movement.
        Aust Orthod J. 1995; 13: 238-248
        • Burstone C.J.
        • Groves M.H.
        Threshold and optimum force values for maxillary tooth movement.
        J Dent Res. 1961; 39: 695-696
        • Andreasen G.F.
        • Zwanziger D.
        A clinical evaluation of the differential force concept as applied to the edgewise bracket.
        Am J Orthod. 1980; 78: 25-40
        • Kapila S.
        • Sachdeva R.
        Mechanical properties and clinical applications of orthodontic wires.
        Am J Orthod Dentofacial Orthop. 1989; 96: 100-109
        • Gale M.S.
        • Darvell B.W.
        Thermal cycling procedures for laboratory testing of dental restorations.
        J Dent. 1999; 27: 89-99
        • Low D.
        • Sumii T.
        • Swain M.V.
        • Ishigami K.
        • Takeda T.
        Instrumented indentation characterization of mouth-guard materials.
        Dent Mater. 2002; 18: 211-215
        • Lendlein A.
        • Langer R.
        Biodegradable, elastic shape-memory polymers for potential biomedical applications.
        Science. 2002; 296: 1673-1676